User Instruction
Test Procedure
The patient should fast four hours prior to the test. Patient should not have taken antibiotics one month prior to the test, or active acid inhibitors five days prior to the test. The patient can normally perform the test without strict supervision. The entire procedure takes about 20 minutes.
 

CHECK
1. Check that the Heliprobe™ instrument is ready to use, by pressing the startkey. The display shall lit up showing the text: "Ready to measure".
   

CAPSULE
2. The patient swallow a capsule, tablet or solution containing 14C-urea with approximately 50 ml of water.
   

WAIT 10 MIN.
3. At the same time, set the timer to target time according to 14C-urea supplier recommendation ("peak" time is typically 10 to 30 minutes).
   

OPEN PACK
4. When the timer rings, open the BreathCard™ (envelope and take out the BreathCard™)
   

BLOW
5. Instruct the patient to exhale gently into the BreathCard™ mouthpiece. The patient may remove the BreathCard™ from his/her mouth in order to speak or breathe normally. Exhaling must continue until the indicator membrane chages from orange to yellow. This takes about 1-4 minutes.
   

INSERT
6. When the indicator has changed colour, the BreathCard™ is ready for analysis and shall be inserted into the Heliprobe™ instrument.
   

START
7. Press the green key on the analyzer and the BreathCard™ analysis will start. The result is automatically analyzed and presented on the display as "Heliprobe™ grade" 0,1 or 2 see below:
  Heliprobe™ 0 : patient not infected
Heliprobe™ 1 : borderline (treatment according to local practice)
Heliprobe™ 2 : patient infected
 
top
User Instruction for Heliprobe™
1. General Description
The Heliprobe™ System 2000 is a diagnostic system for the detection of Helicobacter pylori.
 
1.1 The Front Panel
The front panel consists of one display, two diodes (one red and one green/yellow) and four keys (Print, Start/Stop, Menu, OK). The fuction keys are explained in section 1.2.
The diodes show the status of the Heliprobe™.
  • Constant green light – indicates stand-by-mode. To activate the Heliprobe™ press any key or insert a BreathCard™ and the display will light up.
  • Intermittent green light – the Heliprobe™ is ready for measurement. Insert a Breath Card™ and press the key.
  • Intermittent yellow light – the measurement is in progress. The measurement can be interrupted by pressing the key or withdrawing the BreathCard™.
  • Constant yellow light – the measurement is complete.
  • Confirm the measured values by withdrawing the BreathCard™ or pressing the OK key.
  • Constant red light – an internal system error has occurred. See chapter 4 “Warning and error indication”.
1.2 The Keys
 
PRINTER/COMPUTER ACTIVATE KEY
Transfer Heliprobe™ data to a printer or a computer.
Can also be used to step parameters “down” in the loop-to-loop programming set up mode.
   
START/STOP
Indicates or stops the analyzing if the BreathCard™ is inserted.
   

MENU
OK CONFIRM KEY
By pressing the menu key, a selection of programming possibilities can be selected. When the menus of choice appears on the display, presss the confirm key to enter the program to be changed. Within this program, programming possibilities again can be selected by using the menu and confirm keys. Can also be used to step parameters “up” in the loop-to-loop programming set-up mode.
 
1.3 Power Supply
Connect the power supply (9V DC, 4W) to the lower back side of the Heliprobe™.
 
1.4 Calibration Card
The Calibration Card has the same shapes as the BreathCard™. The Calibration Card must always be inserted in the Heliprobe™ when the instrument is not in use. The Calibration Card is important to ensure correct background radiation level measurement.
 
1.5 Connection of printer or computer
A printer or a computer can be connected at the back of the unit via a RS232 cable (optional).
 
top
2. Operation
2.1 General
The first time the Heliprobe™ is installed, or when the Heliprobe™ is moved to an other location, set-up installation should be performed, in order to adapt the instrument to the actual background radiation environment. This setup installation also enables a verification and/or modification of the programmed parameters for “cut off”, calculations etc.
Note: Do not forget the Calibration Card during the 1000 seconds calibration period after setup, if E-prom has been cleared.
 
2.2 Setup Installation
To activate the set up menu:
- Hold the key while connecting the power supply to the instrument.

The display will show the text:
- Setup menu. Press OK.
- Press “OK.”
(If the “OK” key is not pressed within 40 seconds, the Heliprobe™ will restart automatically).

Note: The loop-to-loop diagram can be run through by pressing the key see (the double edged squares in the diagram). If some function has to be changed or adjusted, select the function by pressing OK as shown in the diagram.

 
2.3 Loop-to-loop Diagram (To view the diagram click)
 
The following keys can also be used as programming keys in the setup programming mode.
  = (Increase)
         
  = (Decrease)
 
2.4 Regular Operation
Incase of short power losses and if the Heliprobe™ has not been moved, the Heliprobe™ can be re-powered without activating the set-up mode. The Heliprobe™ will automatically start a self test showing on the display. After this automatic internal control, the display will show “ready to measure” and the Heliprobe™ is then ready for use.

The Heliprobe™ should always be switched on, with the Calibration Card inserted when not in use.

 
top
3. BreathCard™ Reading
1. Insert the BreathCard™ into the Heliprobe™. The display will show “ready to measure”.
2. Press the key. The standard measurement will last for 250 seconds.
   
3. When the measurement is completed the Heliprobe™ will emit two beeps and the result will be presented on the display as a Heliprobe™ grading 0,1 or 2. By pressing the OK key, the numerical result will be displayed.
 
4. The result will be shown on the display until the measurement is confirmed by pressing the OK key. The result may be printed out by pressing the print key, providing a printer has been connected to the Heliprobe™.
5. By pressing the OK key the display will show the last measurement result at any time.
6. The Heliprobe™ grading 0,1 or 2 is based on clinical trials using 1u Curie 14-C-urea. Using the Heliprobe™ standard setup program the corresponding cut-off values, are shown below.
Heliprobe™ 0 – NOT infected patient. <25
Heliprobe™ 1 – Borderline case. 25< <50
Heliprobe™ 2 – Infected patient, >50
7. Incase that the measurement of the BreathCard™ falls into the Heliprobe™ 1 grade limits, the instrument automatically re-measures the BreathCard™. If the second measurement also falls into Heliprobe™ grade 1, a third measurement is proceeded.
After three measurements the instrument calculates a mean value of all three measurements and a final result is presented. This situation therefore may require up to 750 seconds to be completed. The mean value is also calculated after the first re-measurement.
 
top
4. Warning and error indication
The Heliprobe™ has a number of “indications” implemented in the software to prevent erroneous readings from contamination, etc. The red diode indicated that an error has occurred. The type of warning or error will be displayed, e.g.
  • High background radiation level compared to setup value.
  • Unexpected variation in background radiation during the measurement.
  • Damaged sensor.
Note: For further information contact Noster System AB. See the rear page for address.
 
top
5. Maintenance of the Heliprobe™
The Heliprobe™ may be cleaned using mild cleaning agents. Running water should be avoided since there is high voltage in the Heliprobe™. Avoid insertion of any sharp object into should be carried out by authorized personnel.
 
top
6. Technical Specification
Detector : GM-tubes end window 50 mm in diameter mounted face to face. Internal High voltage, 520V Life time 10E8 pulses
Display : LCD, 16 position x 2 rows, background illumination
Acoustic indication : Beeper
Operating Voltage : 9 V DC (stabilized), External battery eliminator
Power consumption : 4 W
Measuring sensitivity : Activity of 25 Bq 14-C enclosed as during the same condition as in the BreathCard™, give 1 pulse per second
Temperature Range : 0-50 degree C
Humidity : RH max 95%
Weight : 4,2 kg
Serial output : RS 232,
 
Protocol
Baud rate 9 600
No parity
8 data bits
1 stop bit
  Cable available from Noster
The Heliprobe™ system 2000 consist of the following components
Article Article No
Analyzer 2000 A
Power Supply 2000 A-PS
Connector cable 2000 A-CC
BreathCard™ 2000 B
All specifications can be subjected to changes without notice.
 
top
7. Safety Consideration
Pharmaceutical Formula Recommended 14 C-urea doze to be used, is 1µCi, corresponding to a body exposure of 0.0006 mSi. (This exposure is equivalent to one half of an hour in an aeroplane. The total exposure is 250 times less than a mammography).
14C : The carbon isotype “14C” emits low energy b- radiation.
BreathCard :  
Note :
  • The BreathCard™ contains concentrated Alkaline that can irritate the respiratory tract.
  • Only exhale into the BreathCard™. Do not inhale while the BreathCard™ is in mouth.
  • Do not disassemble the BreathCard™.
  • Do not manipulate or ortherwise come into contact with the absorption area or the colour indicator of the BreathCard™.
  The BreathCard™ has a limited shelf-life, see BreathCard™ envelope labeling.
Always put the BreathCard™ back into the envelope before storing or discarding.
The disposal of the BreathCard™ waste must be handled according to local regulations.
 
top
Copy Rights Reserved 2006, Rad-Probe Systems Disclaimer Design by : Ambest Media